Catalog Number 9734298
Device Problems Complete Blockage (1094); Occlusion Within Device (1423)
Patient Problem No Patient Involvement (2645)
Event Date 10/15/2014
Event Type  malfunction  
Manufacturer Narrative
Evaluation of suspect tap as follows: as reported, a guide wire has been broken off inside the tap.Also, the tip of the instrument has been blunted and slightly fluted out.Otherwise, the tap is in good condition and inserts and releases easily from a navlock.
 
Event Description
A medtronic representative received a report from a site that when attempting to clean their 4.5mm cannulated tap, a metal wire became lodged in the cannula.An obstruction, and wire, are stuck in the tap.No further details were provided.There was no patient present when this issue was identified.
 
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Return requested.Replacement 4.5 tap shipped to site (b)(4) 2014.No parts have been returned to manufacturer for analysis.
 
Manufacturer Narrative
Correction: on 13-oct-2015, it was noticed that a previous mdr submission contained incorrect information with regards to the common device name, product code and/or pma/510(k).This mdr is being submitted to correct this information.There is no new information to change the patient information, event description or manufacturer narrative that was previously reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM TAP (CANNULATED)
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC.
826 coal creek circle
louisville CO 80027 9710
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC.
826 coal creek circle
louisville CO 80027 9710
Manufacturer Contact
kristianna bilan
826 coal creek circle
louisville, CO 80027-9710
7208902338
MDR Report Key4243376
MDR Text Key19013939
Report Number1723170-2014-01210
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K124004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 11/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9734298
Device Lot Number130722
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/16/2014
Initial Date FDA Received11/11/2014
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received01/14/2015
11/10/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-