Catalog Number 9734298 |
Device Problems
Complete Blockage (1094); Occlusion Within Device (1423)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/15/2014 |
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation of suspect tap as follows: as reported, a guide wire has been broken off inside the tap.Also, the tip of the instrument has been blunted and slightly fluted out.Otherwise, the tap is in good condition and inserts and releases easily from a navlock.
|
|
Event Description
|
A medtronic representative received a report from a site that when attempting to clean their 4.5mm cannulated tap, a metal wire became lodged in the cannula.An obstruction, and wire, are stuck in the tap.No further details were provided.There was no patient present when this issue was identified.
|
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.Return requested.Replacement 4.5 tap shipped to site (b)(4) 2014.No parts have been returned to manufacturer for analysis.
|
|
Manufacturer Narrative
|
Correction: on 13-oct-2015, it was noticed that a previous mdr submission contained incorrect information with regards to the common device name, product code and/or pma/510(k).This mdr is being submitted to correct this information.There is no new information to change the patient information, event description or manufacturer narrative that was previously reported.
|
|