Model Number 3003-009
Device Problem Difficult or Delayed Positioning (1157)
Patient Problem Laceration(s) (1946)
Event Date 06/26/2014
Event Type  malfunction  
Event Description
The customer reported that the pt's head slipped off the headrest while pinning the pt.Pt was in prone position.Laceration occured, but no serious pt injury.
 
Manufacturer Narrative
Failure reported by customer but product is within specs.Device evaluated and alleged failure could not be duplicated.Corrective action: instructed customer to properly position the pt's head.Ensure that the 2 skull pins in the rocker arm are equidistant from the centerline of the head and that all 3 pins are penetrating the skull equally.The angle of the pins should be as close to 90 degrees possible to the pt's skull.Test that the skull clamp ext.Arm is fully engaged in the clamp base.Lock the index knob so that the arrow and "lock" are aligned.Returned product for use and sent user another copy of the instruction manual.Suspected root cause: wrong pinning technique.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO TEFLON SKULL CLAMP
Type of Device
HEADHOLDER
Manufacturer (Section D)
PROMED INSTRUMENTS GMBH
boetzinger strasse 38
freiburg 7911 1
GM  79111
Manufacturer Contact
boetzinger strasse 38
freiburg 79111
6138422242
MDR Report Key4249897
MDR Text Key5106739
Report Number3003923584-2014-00017
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001808
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/22/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3003-009
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/08/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/08/2014
Initial Date FDA Received11/06/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-