• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - SAN JOSE INTELLAMAP ORION?; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - SAN JOSE INTELLAMAP ORION?; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number M004RC64S0
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/08/2014
Event Type  malfunction  
Event Description
Reportable based on analysis completed on 11nov2014.It was reported that during an ablation procedure in the right atrium, the catheter was unable to be deployed.The catheter was removed and they found that there was a feature on the side of the handle that had became loose.The physician was able to complete the procedure using this device.Patient status reported as fine.Returned device analysis revealed that the safety button was completely detached.
 
Manufacturer Narrative
(b)(4).Visual inspection of the returned device revealed that the device safety button was loose and detached.The handle was opened and the safety button was found completely detached.There was evidence of adhesive in two points inside the component, however it is broken.The device was able to perform deploy and undeploy without issues during the dimensional test.The handle and safety button sections were sent to mtac for additional analysis to determine the failure mode.This analysis revealed adhesive failure sites presented evidence of cleavage/peel mechanical overload.In addition to adhesive failure, the safety button presented delamination and particle rich in sodium and chloride.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLAMAP ORION?
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC - SAN JOSE
150 baytech drive
san jose CA 95134
Manufacturer (Section G)
BOSTON SCIENTIFIC - SAN JOSE
150 baytech drive
san jose CA 95134
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key4251261
MDR Text Key18265837
Report Number2134265-2014-07369
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122461
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2015
Device Model NumberM004RC64S0
Device Catalogue NumberRC64S
Device Lot Number17088826
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/11/2014
Initial Date FDA Received11/14/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/03/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-