BOSTON SCIENTIFIC - SAN JOSE INTELLAMAP ORION?; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number M004RC64S0 |
Device Problem
Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/08/2014 |
Event Type
malfunction
|
Event Description
|
Reportable based on analysis completed on 11nov2014.It was reported that during an ablation procedure in the right atrium, the catheter was unable to be deployed.The catheter was removed and they found that there was a feature on the side of the handle that had became loose.The physician was able to complete the procedure using this device.Patient status reported as fine.Returned device analysis revealed that the safety button was completely detached.
|
|
Manufacturer Narrative
|
(b)(4).Visual inspection of the returned device revealed that the device safety button was loose and detached.The handle was opened and the safety button was found completely detached.There was evidence of adhesive in two points inside the component, however it is broken.The device was able to perform deploy and undeploy without issues during the dimensional test.The handle and safety button sections were sent to mtac for additional analysis to determine the failure mode.This analysis revealed adhesive failure sites presented evidence of cleavage/peel mechanical overload.In addition to adhesive failure, the safety button presented delamination and particle rich in sodium and chloride.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined (b)(4).
|
|