• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES OBERDORF DRILL/BURR ATTACHMENT FOR 2.35MM SHAFTS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES OBERDORF DRILL/BURR ATTACHMENT FOR 2.35MM SHAFTS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 05.001.123
Device Problem Device Difficult to Setup or Prepare (1487)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/28/2014
Event Type  malfunction  
Event Description
It was reported that during an unspecified surgical procedure, it was observed that the attachment device could not load burrs.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was duplicated and confirmed.The assignable root cause was determined to be most likely due to worn locking mechanism caused by exposure to organic debris and materials which led to corrosion and normal component wearout.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL/BURR ATTACHMENT FOR 2.35MM SHAFTS
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH44 36
SZ  CH4436
Manufacturer (Section G)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH44 36
SZ   CH4436
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key4252089
MDR Text Key5106784
Report Number8030965-2014-10647
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK043310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/29/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.123
Device Lot Number1659
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/03/2015
Initial Date FDA Received11/14/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/08/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
BURR DEVICE
-
-