• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problems No Display/Image (1183); Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problems No Patient Involvement (2645); No Patient involvement (2648)
Event Date 10/13/2014
Event Type  malfunction  
Event Description
Sorin group (b)(4) received a report that the touch screen of th s5 roller pump was unresponsive during set up.There was no patient involvement.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that stated at turning on the s5 during setup, the touch screen was off, but the user could control the pump with the knob and the leds were in normal operation.Upon checking the internal part of the pump, the service representative found the connector cable loose.The cable has possible compression damage by the structure.A new cable was ordered for the customer.There was no patient involvement.No further investigation is required as no trend increase for this kind of issue was identified.No nonconformities were noted during device history record review.The issue will be monitored for trends and if identified, corrections will be recommended.
 
Event Description
Sorin group (b)(4) received a report that the touch screen of the s5 roller pump was unresponsive during set up.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
cheri voorhees
14401 w. 65th way
arvada, CO 80004
3034676527
MDR Report Key4259096
MDR Text Key5110805
Report Number1718850-2014-00425
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeCI
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 08/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/15/2014
Initial Date FDA Received11/12/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/04/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-