It was reported to medtronic neurosurgery that the device was used on the patient following a lumbar drain procedure.According to the report, on (b)(6) 2014, in the afternoon, the staff nurse on duty noticed that there was no csf collected in the chamber.Reportedly, at around 6:60 pm it was found that there still wasn¿t any csf collected.According to the report, it was noticed that there was yellowish fluid on the floor just below the duet system and it was discovered that there was a disconnection between the drip chamber and the 3-way stopcock.The report stated that there was no injury to the patient and that the patient is in a stable condition.
|
The device was patent.The stopcock below the drip chamber was disconnected from the drip chamber.The damage led to the device not meeting the requirements for the leakage test.There was adhesive noted at the connection.The drainage bag was stuck to the connector and could not be unscrewed.Application of excessive force in the tightening or loosening of the drainage bag connection may have led to the damage.The instructions for use (ifu) that accompanies the device caution that ¿in order to avoid possible cracking of the luer connectors after cleaning with alcohol, or a disinfectant containing alcohol allow to air dry completely prior to connecting the system¿.Also, the ifu cautions that ¿all connections should be finger tightened.Over tightening can cause cracks and leaks to occur¿.A review of the manufacturing records showed no anomalies.(b)(4).
|