• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO SYSTEM 5 DUAL TRIGGER ROTARY HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO SYSTEM 5 DUAL TRIGGER ROTARY HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 4205000000
Device Problem Unintended Power Up (1162)
Patient Problem No Patient Involvement (2645)
Event Date 10/23/2014
Event Type  malfunction  
Event Description
It was reported that during testing conducted at the manufacturer facility, the device ran without user activation after the trigger was released.No medical intervention and no adverse consequences were reported with this event.As this event occurred during testing at the manufacturer facility, there was no patient involvement and no delay to a surgical procedure.
 
Manufacturer Narrative
Upon disassembly, a displaced straight pin was found, which can lead to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYSTEM 5 DUAL TRIGGER ROTARY HANDPIECE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key4263839
MDR Text Key21447602
Report Number0001811755-2014-04184
Device Sequence Number1
Product Code KIJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K972367
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number4205000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2014
Initial Date Manufacturer Received 10/23/2014
Initial Date FDA Received11/20/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-