Procedure: sleeve gastrectomy.According to the reporter: during fire sequence, there was a popping noise in shaft and stopped mid-fire, just half way through the staple line.After removing device, shaft would only articulate one to the left, but when articulated to right, motor inside idrive ultra would just run without any action from the reload.The egiaadapt was put onto the device to determine if it were the handle or adapter, it was concluded that egiaadaptxl was the issue, while short handle worked fine.List any other products used with device: idrvultra1 was the device used in patient: yes was there patient injury/hazard: no there was no unanticipated tissue loss, no unanticipated extension of incision more than 1 inch, no unanticipated blood loss of more than 500cc and no delay over 30 minutes.No device fell in patients cavity and no device fragment left in patient.
|