MAUDE Adverse Event Report: DEPUY MITEK RIGIDFIX PLA 3.3 MM FEM/TIB ST CROSS PIN KIT, CONTAINS 210133 & 210333; MITEK ACL IMPLANTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK RIGIDFIX PLA 3.3 MM FEM/TIB ST CROSS PIN KIT, CONTAINS 210133 & 210333; MITEK ACL IMPLANTS Back to Search Results
Catalog Number 210133
Device Problem Unknown (for use when the device problem is not known) (2204)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/28/2014
Event Type  Injury  
Event Description
The sales rep reported that during an acl repair that the pin guides of customer's rigidfix st 3.3 cross pin kit and rigidfix st guideframe would bind together.The surgeon replaced the cross pin kit 2 times with the same issue.The sales rep reported that the guideframe was found to be warped and bent.The sales rep reported that no metal shavings were created.The surgeon completed the procedure with a 4th rigidfix st 3.3 cross pin kit and a second same type guideframe with no patient consequences.There was a 40 minute delay in the procedure.The sales rep could not provide the lot number or age for the guideframe but reported it has seen heavy use.See associated med watch 1221934-2014-00519, 1221934-2014-00520, 1221934-2014-00521.
 
Manufacturer Narrative
The complaint device is not being returned and therefore a physical evaluation cannot be performed.From the event details provided, the failure of the devices in this complaint could not be specifically determined and a root cause cannot be discerned.A batch record review has been conducted and the results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaints for this lot of devices that were released to distribution.Therefore, at this point in time, based on the complaint history, no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIGIDFIX PLA 3.3 MM FEM/TIB ST CROSS PIN KIT, CONTAINS 210133 & 210333
Type of Device
MITEK ACL IMPLANTS
Manufacturer (Section D)
DEPUY MITEK
325 paramount drive
quality department
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK
325 paramount drive
na
raynham MA 02767
Manufacturer Contact
siri belur
325 paramount drive
quality department
raynham, MA 02767
8003824682
MDR Report Key4271389
MDR Text Key18643314
Report Number1221934-2014-00518
Device Sequence Number1
Product Code HTY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K974341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 10/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2017
Device Catalogue Number210133
Device Lot Number3784003
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date10/28/2014
Device Age3 MO
Event Location Hospital
Date Report to Manufacturer10/28/2014
Initial Date Manufacturer Received 10/28/2014
Initial Date FDA Received11/21/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/09/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-