• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRIS INTERNATIONAL ICHEM VELOCITY; AUTOMATED URINE CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRIS INTERNATIONAL ICHEM VELOCITY; AUTOMATED URINE CHEMISTRY ANALYZER Back to Search Results
Catalog Number 700-7177-001
Device Problems Charred (1086); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/22/2013
Event Type  malfunction  
Event Description
An iris field service engineer (fse) observed the doser fluid system (dfs) z axis motor flex cable was partially disconnected and melted on the flex cable.
 
Manufacturer Narrative
During service visit on ichem velocity, fse observed the doser fluid system 2 axis motor flex cable was partially disconnected and melted the flex cable.Fse replaced the board and flex cable and calibrated the cgm.Fse completed velocity performance verification checklist and observed pt samples running normally.System was operational.There were no reports of injuries, visible flames, smoke emission, and the fire department was not called.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICHEM VELOCITY
Type of Device
AUTOMATED URINE CHEMISTRY ANALYZER
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave.
chatsworth CA 91311
Manufacturer Contact
sudha gupta
9172 eton ave.
chatsworth, CA 91311
8185277272
MDR Report Key4272831
MDR Text Key21447105
Report Number2023446-2014-00163
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/22/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number700-7177-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/22/2013
Initial Date FDA Received10/08/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-