IRIS INTERNATIONAL ICHEM VELOCITY; AUTOMATED URINE CHEMISTRY ANALYZER Back to Search Results
Catalog Number 700-7177-001
Device Problems False Negative Result (1225); Failure to Prime (1492); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/19/2014
Event Type  malfunction  
Event Description
Customer reported issues with false negative blood reporting.
 
Manufacturer Narrative
Ichem velocity urine chemistry reported false negative blood for pt samples.Two samples were reported out to the physician.It is unk if pt management was changed as a result.Fse noticed the analyzer was becoming de-primed during normal operation, leading to air bubbles in tubing.Fluid block assembly was replaced and instrument stayed primed.Fse completed velocity performance verification checklist and system was operational.Upon further investigation, 2 pt samples on urine chemistry reported false negative for blood and these results were reported out to the physician.It is unk at the present time if there was a change in pt management as a result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICHEM VELOCITY
Type of Device
AUTOMATED URINE CHEMISTRY ANALYZER
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave.
chatsworth CA 91311
Manufacturer Contact
sudha gupta
9172 eton ave.
chatsworth, CA 91311
8185277272
MDR Report Key4272840
MDR Text Key5228945
Report Number2023446-2014-00157
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number700-7177-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/19/2014
Initial Date FDA Received10/08/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown