Same case as: 2134265-2014-07058.Reportable based on device analysis completed on 28oct2014.It was reported that the catheter could not be loaded on the wire.A 1.75mm rotalink¿ plus and a 330cm rotawire¿ and wireclip¿ torquer were selected to treat the target lesion.However, it was not possible to load the rotalink onto the rotawire; it seemed the the burr did not have an end hole.The device did not go inside the patient.No patient complications were reported.However, device analysis revealed a piece of guidewire inside the annulus of the burr.
|