This is a report of a home patient who experienced a connection issue between the heater bag and the cassette.This occurred during dwell 1 of 9 in peritoneal dialysis therapy.The caregiver stated that when they connected the bag, it felt like it would not tighten and it kept spinning.The technical service representative assisted the caller in ending therapy and removing the supplies.The caregiver would start over with all new supplies.There was no patient injury, medical intervention, or adverse event associated with this report.No additional information is available at this time.
|
(b)(4).The sample was received and an evaluation of the device was performed.Visual inspection, leak testing, device-device interaction testing, clamp function test, and clear passage testing were performed with no issues noted.Sample analysis determined the product met all specifications.Should relevant additional information become available, a supplemental report will be submitted.
|