Model Number 70104.0852
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/23/2013
Event Type  malfunction  
Event Description
Blood monitoring unit (bmu) 40 with serial number (b)(4), at (b)(6) hospital, venous probe failed to calibrate/initialized.The customer held the probe in the slot in order for the unit to initialize.Reference: complaint (b)(4).
 
Manufacturer Narrative
(b)(4).On (b)(6) 2013, additional information was received that the probe was manually held in the slot to calibrate.Also the user rebooted the device one time.Holding the probe is not in accordance with the user manual.The probe must be fixed in the slot.Therefore, the unit did not malfunction.The unit function as intended by not initializing because the probe was not affixed to the slot.Reference: complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOOD MONITORING UNIT (BMU) 40
Type of Device
BLOOD GAS MONITOR
Manufacturer (Section D)
MAQUET CARDIOPULMOARY AG
kehler strasse 31
rastatt, 7643 7
GM  76437
Manufacturer (Section G)
MAQUET CARDIOPULMOARY AG
kehler strasse 31
rastatt, 7643 7
GM   76437
Manufacturer Contact
tina evancho
45 barbour pond drive
wayne, NJ 07470
9737097265
MDR Report Key4280523
MDR Text Key5300110
Report Number8010762-2014-00743
Device Sequence Number1
Product Code DRY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090147
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/24/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Model Number70104.0852
Device Catalogue Number70104.0852
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date01/24/2013
Event Location Hospital
Date Report to Manufacturer01/24/2013
Initial Date Manufacturer Received 01/24/2013
Initial Date FDA Received11/26/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-