MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3058
Device Problems High impedance (1291); Unable to Obtain Readings (1516)
Patient Problem Scar Tissue (2060)
Event Type  Injury  
Event Description
It was reported that there had been ¿impedance issues¿ on their implantable neurostimulator (ins) since 2011.It was noted that the patient was having a revision and was currently in the or at the time of report.The reason for the revision was that the patient¿ ins system had impedances >4000 ohms specifically on electrode pairs with electrode #0.Further impedance testing showed that a couple of electrode pairs showed ¿???¿ after increasing the pulse width (pw) to 240.It was then advised to also increase the voltage to 2.0 volts during the impedance testing which then resolved the ¿???¿ issue.It was also noted that the healthcare professional used the left side lead that had a lot of scar tissue near it.There were no further details, interventions or outcome reported regarding this event.Further follow-up is being conducted to obtain this information.If additional information is received, a follow-up report will be sent.
 
Manufacturer Narrative
Product id 3889-28, lot# v732559, implanted: 2011 (b)(6); product type lead product id 3037, serial# (b)(4); product type programmer, patient.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key4281706
MDR Text Key5230673
Report Number3004209178-2014-22254
Device Sequence Number1
Product Code EZW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 11/06/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2013
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/06/2014
Initial Date FDA Received11/26/2014
Date Device Manufactured10/31/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age00006 YR
-
-