A discordant, falsely elevated troponin result was obtained upon repeat testing on one patient sample on an advia centaur cp instrument.The discordant result was not reported to the physician(s).The sample initially resulted as expected when tested on the same instrument and falsely elevated upon first repeat.The sample was repeated a second time on the same instrument, matching the result of the initial run.It is unknown if the corrected result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated troponin result.
|
A siemens customer service engineer (cse) was dispatched to the customer site.Prior to cse arrival, precision testing was run on a negative patient pool.Out of the sixty samples tested, one resulted near the customer's cutoff for a positive troponin value.After evaluation of the instrument and instrument data, the cse replaced the valve manifold sample syringe and tubing, reagent probe, and a defected ring position sensor.The cse calibrated the incubation ring.The cause of the discordant, falsely elevated troponin result is unknown.Siemens is investigating this issue.
|
The initial mdr 2432235-2014-00662 was filed on november 26, 2014.Additional information (12/16/2014): the issue has been resolved with the siemens customer service engineer's (cse) last visit.The cause of the discordant, falsely elevated troponin result is unknown.The details of the last service visit were provided in the initial mdr.The instrument is performing according to specifications.No further evaluation of the device is required.
|