• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG QUART ARTERIAL FILTER, MODEL HBF 140; FILTER, BLOOD, CPB, ARTERIAL LINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG QUART ARTERIAL FILTER, MODEL HBF 140; FILTER, BLOOD, CPB, ARTERIAL LINE Back to Search Results
Model Number HBF-140 U-J
Device Problems Particulates (1451); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 09/05/2012
Event Type  malfunction  
Event Description
On (b)(6) 2012, at (b)(6) it was reported that a red particle was found in the arterial filter quart hbf 140-j.There was no patient involvement.(b)(4).
 
Manufacturer Narrative
(b)(4).The foreign particulate matter originated from an abrasion of the red stopcock.Manufacturing identified the root cause and polished the molding tool to cease abrasion.A diameter reduction of the stopcock holder has been implemented to cease further red particulate matter.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUART ARTERIAL FILTER, MODEL HBF 140
Type of Device
FILTER, BLOOD, CPB, ARTERIAL LINE
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
kehler strasse 31
rastatt, 7643 7
GM  76437
Manufacturer (Section G)
MAQUET CARDIOPULMONARY AG
kehler strasse 31
rastatt, 7643 7
GM   76437
Manufacturer Contact
tina evancho
maquet medical system, usa
45 barbour pond drive
wayne, NJ 07470
9737097265
MDR Report Key4282881
MDR Text Key5299612
Report Number8010762-2014-00473
Device Sequence Number1
Product Code DTM
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K001787
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/04/2012
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/23/2013
Device Model NumberHBF-140 U-J
Device Catalogue Number70103.1618
Device Lot Number70065655
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2012
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/04/2012
Device Age18 MO
Event Location Hospital
Date Report to Manufacturer10/04/2012
Initial Date Manufacturer Received 10/04/2012
Initial Date FDA Received11/27/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/04/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-