(b)(4) it was reported that a patient underwent an atrial fibrillation procedure with a lasso nav duo loop eco catheter and at the end of the procedure there was difficulty retracting the lasso catheter from the sl1 sheath, therefore it was removed from patient.Upon removal it was noted that the catheter coating detached exposing wires with no missing pieces.This complaint is reportable because inner components are exposed and there is a higher risk that this could potentially contribute to a death or serious injury.The bwi failure analysis lab received the device for evaluation.Upon receiving, it was found that the spine cover was ruptured and scrunched towards the catheter¿s tip.A white material was also found stuck underneath electrical ring #18 at the proximal end.Upon investigation, the white material was determined to come from the sheath used.Additionally, a scanning electron microscope analysis was carried out and the results showed that the distal section of the rupture presented evidence of elongations.Proximal section of the rupture presented evidence of physical damages, elongations and accordion condition.These conditions suggest that the device was induced to an excessive application of tension with the sheath on the rupture area.The accordion condition also suggests that the tension applied to the device was on the direction to the distal tip.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.All the catheters are inspected for visual damages before packaging.On line inspections are in place to prevent this type of damage/defect from leaving the facility.The customer complaint has been verified.An internal corrective action has been opened to address the accordion condition for the lasso nav duo loop eco family.
|