Model Number H749CBM3250150 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/07/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that a shaft break occurred.A 15/2.50 flextome® cutting balloon® monorail was used to treat the target lesion.During procedure, it was noted that the shaft broke.There were no patient complications reported.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: the device was returned for analysis.A visual examination of the returned device confirmed a hypotube shaft break at 2.5mm from the catheter strain relief.The hypotube was also kinked at several locations along its length.The inner was kinked at 21mm proximal to the distal tip and the balloon was also kinked in this area.This type of damage is consistent with excessive force being applied to the delivery system.No damage was noted to the tip or blades of the device.The batch number is unknown and the manufacturing records for the complaint device could not be reviewed.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
|
Event Description
|
It was reported that a shaft break occurred.A 15/2.50 flextome cutting balloon monorail was used to treat the target lesion.During procedure, it was noted that the shaft broke.There were no patient complications reported.
|
|
Search Alerts/Recalls
|