• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR XP; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR XP; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number ADVIA CENTAUR XP
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/10/2014
Event Type  malfunction  
Event Description
A discordant, falsely elevated troponin result was obtained on one patient sample on an advia centaur xp instrument.The discordant result was not reported to the physician(s).The sample was repeated on the same instrument and an alternate instrument, resulting lower.The corrected result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated troponin result.
 
Manufacturer Narrative
A siemens customer service engineer (cse) was dispatched to the customer site.After evaluation of the instrument and instrument data, the cse replaced the wash-blocks and probe guides due to liquid splatter and crossover, all aspiration valve tubing due to poor aspiration, and tube connectors which were found leaking.The cse ran calibration and quality controls (qc), with poor results.The cse discovered leaks from the base pump and replaced the connectors.Calibrations and qc were rerun, resulting within range.The cause of the discordant, falsely elevated troponin results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR XP
Type of Device
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS
manufacturing limited
registration number: 8020888
chapel lane, swords, co., dublin
EI  
Manufacturer Contact
cassandra kocsis
511 benedict avenue
tarrytown 10591
9145242687
MDR Report Key4287921
MDR Text Key5085499
Report Number2432235-2014-00700
Device Sequence Number1
Product Code MOI
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K971418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 11/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberADVIA CENTAUR XP
Device Catalogue Number078-A010-02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/10/2014
Initial Date FDA Received12/01/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/14/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-