• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® RESUSCITATION SYSTEM MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® RESUSCITATION SYSTEM MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Crack (1135); Device Damaged Prior to Use (2284); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/28/2014
Event Type  malfunction  
Event Description
It was initially reported that the when the autopulse platform was removed from the shipping carton, it was noticed that the base covering was cracked at the mounting location of the top handle area.No other issues were reported and no further information was provided.The autopulse platform was subsequently returned to zoll for investigation.During review of the platform's archive data, it was observed that a user advisory (ua) 20 (position out of range) and a (ua) 45 (not at "home" position after power-on/restart) message occurred on the reported event date of (b)(6) 2014.Although the customer did not report these, ua 20 and ua 45 are considered reportable malfunctions.
 
Manufacturer Narrative
The autopulse platform (s/n (b)(4)) was returned to the manufacturer for analysis.The customer reported that the cover was cracked upon receipt.Further investigation reveals that the damage appears to have been due to mishandling.A review of the autopulse archive was performed and the archive data shows ua20 (position out of range) and ua45 (not at "home" position after power-on/restart) faults occurred on the reported event date of (b)(6) 2014.However, these faults are not related to the reported complaint.Functional testing was performed and the platform passed testing.Based on the investigation, the part identified for replacement was the front cover.In summary, the reported complaint was confirmed during visual inspection.The fault was found to be due to mishandling.The root cause for ua20 and ua45 faults could not be determined.Upon replacement of the front cover, the platform was re-evaluated through functional testing and the platform passed all testing criteria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION SYSTEM MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
joy patel
2000 ringwood ave,
san jose, CA 95131
4084192957
MDR Report Key4292497
MDR Text Key5272652
Report Number3010617000-2014-00642
Device Sequence Number1
Product Code DRM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 11/20/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/20/2014
Initial Date FDA Received12/02/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-