Model Number IMPAX CV 2.08
Device Problem Installation-Related Problem (2965)
Patient Problem No Information (3190)
Event Date 11/06/2014
Event Type  malfunction  
Event Description
Agfa submitted mdr report #1225058-2010-00001 to the fda on june 7, 2010 for a site in the us.A (b)(6) occurrence is being reported for the same issue/same device: impax cv results management administration tool (rmat).This is an internal discovery determined during the implementation of the associated problem correction plan, rmat verification, as reported in fda z-2112-10.(b)(6).
 
Manufacturer Narrative
An agfa clinical analyst performed a retro-analysis and reported the findings to agfa service and agfa product quality manager.Agfa's investigation into this occurrence of rmat customization has revealed that this specific change had the potential to introduce clinical inaccuracies in patient reports.Specifically: gui type changed which could lead to contradictory statements.Bio prosthetic function finding group changed from radio to box, allowing for contradictory statements to appear on the report.Reports with inappropriate use of gui type mitral valve prosthetic mechanical prosthetic function.Potential impact: contradictory statements can now be made based on findings in the group.Report presented downstream is different from what the reading physician intended.There has been no reported patient harm for this occurrence.A reportable correction is underway for this issue and has been reported to the fda via reference #z-2112-10.Agfa will follow the rmat post market verification work instructions to correct the sentence finding.Any further investigation for the site described in this report will be documented in the ongoing cfr part 806 reporting.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPAX CV OUTBOUND (RESULTS MANAGEMENT 2.08)
Type of Device
PICTURE ARCHIVING AND COMMUNICATION
Manufacturer (Section D)
AGFA HEALTHCARE CORP.
1 crosswind rd.
westerly RI 02891
Manufacturer Contact
deborah huff
10 south academy street
greenville, SC 29601
8644211754
MDR Report Key4293149
MDR Text Key5079048
Report Number1225058-2014-06266
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050228
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 11/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberIMPAX CV 2.08
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/06/2014
Initial Date FDA Received11/21/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2002
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-