• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGFA HEALTHCARE CORP IMPAX CV OUTBOUND (RESULTS MANAGEMENT 2.08); PICTURE ARCHIVING AND COMMUNCATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGFA HEALTHCARE CORP IMPAX CV OUTBOUND (RESULTS MANAGEMENT 2.08); PICTURE ARCHIVING AND COMMUNCATION Back to Search Results
Model Number IMPAX CV 2.08
Device Problem Installation-Related Problem (2965)
Patient Problem No Information (3190)
Event Date 10/31/2014
Event Type  malfunction  
Event Description
Agfa submitted mdr report# 1225058-2010-00001 to the fda on june 7, 2010 for a site in the us.A (b)(6) occurrence is being reported for the same issue/same device: impax cv results management administration tool (rmat).This is an internal discovery determined during the implementation of the associated problem correction plan, rmat verification, as reported in fda z-2112-10.
 
Manufacturer Narrative
An agfa clinical analyst performed a retro-analysis and reported the findings to agfa service and agfa product quality manager.Agfa's investigation into this occurrence of rmat customizations has revealed that this specific change had the potential to introduce clinical inaccuracies in pt reports.Specifically: the original sentence finding of "no tricuspid calcification is visualized" has been changed to:" there is a ruptured chordae involving the anterior tricsupid valve leaflet".In this issue, the original sentence states the tricuspid leaflets are without vegetation.The sentence change however, contains clinically different content indicating the presence of anterior tv leaflet chordae rupture.This could lead to incorrect diagnosis and or inappropriate treatment as a chordae rupture would indicate the presence of tricuspid regurgitation and may indicate the need for tv leaflet repair.If this sentence change is found in the production environment, the ca believes that the potential for misdiagnosis is greater than the potential for inappropriate treatment as the referring surgeon would assess the images prior to intervention.Potential impact: if the change was made after reports had been generated with the original sentence selected, those reports prior to the change could display incorrect clinical data if viewed in rm or if used as a template for a repeat study on that pt.As soon as agfa is made aware of any affected reports related to this specific finding, follow-up mdrs will be submitted for each associated study date and respective medical record number/s (mrn) identified.There has been no reported pt harm for this occurrence.A reportable correction is underway for this issue and has been reported to the fda via reference # z-2112-10.Agfa will follow the rmat post market verification work instructions to correct the sentence finding.Any further investigation for the site described in this report will be document in the ongoing cfr part 806 reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPAX CV OUTBOUND (RESULTS MANAGEMENT 2.08)
Type of Device
PICTURE ARCHIVING AND COMMUNCATION
Manufacturer (Section D)
AGFA HEALTHCARE CORP
1 crosswind rd.
westerly RI 02891
Manufacturer Contact
deborah huff
10 south academy street
greenville, SC 29601
8644211754
MDR Report Key4293353
MDR Text Key5083602
Report Number1225058-2014-06320
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050228
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberIMPAX CV 2.08
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/31/2014
Initial Date FDA Received11/24/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2002
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-