• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY VISIAN ICL (IMPLANTABLE COLLAMER LENS); INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY VISIAN ICL (IMPLANTABLE COLLAMER LENS); INTRAOCULAR LENS Back to Search Results
Model Number MICL13.2
Device Problem Device Slipped (1584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/06/2014
Event Type  malfunction  
Event Description
The reporter stated the surgeon inserted a 13.2mm micl13.2 implantable collamer lens.The surgeon attempted to open the lens in the patient's eye and the lens slipped.The lens was removed within the same surgery and another icl was implanted with no problem.There was no patient injury.
 
Manufacturer Narrative
(b)(4): evaluation: method - work order search.Results - a lens work order search was performed and no similar complaints were found within the work order.Visual inspection of the returned product found the lens optic and one haptic torn.The lens was returned in liquid.Conclusions - (no conclusion can be drawn): based on the complaint history, work order search and the evaluation of the returned product, a specific root cause of the event could not be determined.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIAN ICL (IMPLANTABLE COLLAMER LENS)
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
althea watson
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key4295759
MDR Text Key5052344
Report Number2023826-2014-01043
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 11/06/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date01/31/2016
Device Model NumberMICL13.2
Other Device ID NumberDIOPTER -11.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/06/2014
Initial Date FDA Received12/03/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/10/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTRIDGE MODEL AND LOT NUMBER UNK; INJECTOR MODEL AND LOT NUMBER UNK
-
-