Result: the ruby coil pusher assembly was kinked approximately 5.0 cm from the proximal end.The introducer sheath friction lock was distal of the mid-joint.There was a gap between the coil and the distal detachment tip (ddt).The ruby coil pusher assembly pet lock was intact on the proximal end of the pusher.Conclusion: the complaint has been evaluated.The complaint indicates that there was resistance and trouble getting the coil through a guide catheter.During the evaluation of the coil, it appears that the ruby coil 28x60 was slightly stretched and the proximal end of the pusher assembly hypo-tube was kinked.The cause of the resistance described by the physician was unknown.The coil could not be advance out of the introducer sheath into a penumbra microcatheter to see if the issue in the complaint could be recreated.It is also not known if the guide catheter that was being used was damaged.The kink in the proximal end of the pusher assembly and damage to the coil was likely due to the physician attempting to push the coil against resistance in the guide catheter.The cause of this complaint cannot be determined.These devices are 100% functional tested during process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|