Result: the penumbra coil 400 pusher assembly was fractured approximately 5.0 cm from the proximal end where the hypo-tube tab was located.The pusher assembly was slightly kinked approximately 20.5 cm from the proximal end.The coil was intact with the pusher assembly.The distal tip of the introducer sheath was ovalized.The distal tip of the coil was kinked.Conclusion: the complaint has been evaluated.The complaint indicates that the penumbra coil 400 was noticed by the physician as defective.Evaluation of the returned product was confirmed.The proximal end of the penumbra coil 400 pusher assembly was fractured.This type of damage typically occurs if the device was improperly handled during the removal from the packaging hoop or preparation.The distal end of the penumbra coil 400 introducer sheath was ovalized, and the distal tip of the coil was kinked; however, this damage was not mentioned in the complaint.These devices are 100% functional tested during process inspection visually inspected for damage.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|