• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS OSCILL-SAW-ATTACHM W/QUICK-COUPL F/COMPA; BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS OSCILL-SAW-ATTACHM W/QUICK-COUPL F/COMPA; BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 511.801
Device Problems Connection Problem (2900); Noise, Audible (3273)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Event Description
It was reported from (b)(6) that during pre-testing, before surgery, it was discovered that the attachment device noisy while in use and could not secure the saw blade as intended.As a result, the device did not pass the pre-test.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There was no patient involvement.There were no injuries, medical intervention or prolonged hospitalization reported.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that there was pitting in the housing and the bearing was worn.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If information is obtained that was not available, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSCILL-SAW-ATTACHM W/QUICK-COUPL F/COMPA
Type of Device
BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH44 37
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH44 37
SZ   CH4437
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key4310214
MDR Text Key16633997
Report Number3009450871-2014-10750
Device Sequence Number1
Product Code GFA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511.801
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/17/2015
Initial Date FDA Received12/09/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received08/03/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SAW BLADE DEVICE
-
-