Model Number 8374
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
The customer reported when the alarm is selected to voice and tone, the voice portion does not sound.Customer did not provide the date of event.No pt incident or injury was reported.
 
Manufacturer Narrative
Results: eval confirmed the reported issue; the pre-recorded voice message does not play, only clicking sounds is heard due to a non-functioning voice chip.The unit's tone plays properly.There are scuff marks on the back of the enclosure.No other physical damage is observed.Note: all complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.No corrective or preventative actions are necessary at this time.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KEEPSAFE DELUXE
Manufacturer (Section D)
J. T. POSEY CO.
arcadia CA
Manufacturer (Section G)
J. T. POSEY CO.
5635 peck rd.
arcadia CA 91006
Manufacturer Contact
pam wampler, product surveill
5635 peck rd.
arcadia, CA 91006
6264433134
MDR Report Key4317386
MDR Text Key15114874
Report Number2020362-2014-00372
Device Sequence Number1
Product Code KMI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/22/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8374
Device Catalogue Number8374
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer11/03/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/22/2014
Initial Date FDA Received11/21/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
OVER-THE-MATTRESS SENSOR PAD MODEL:8307 LOT#UNK
-
-