Photographs of the affected unit were provided for evaluation.Review of the returned photos confirmed the catheter had broken.A review of the device history records revealed no irregularities during the manufacture of reported lot number 3331062.No issues affecting the radiopacity of the device in question were observed during this review.Conclusions: based on the provided photos, it is concluded that the catheter may have come into contact with some sort of sharp object or instrument during use, resulting in the break.
|