• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC.; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC.; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ATS45
Device Problems Break (1069); Crack (1135)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/13/2014
Event Type  malfunction  
Event Description
It was reported that during a lavh procedure, the device broke, snapped and cracked when firing.A second device was used to complete the case with no patient consequence.There is no further information available.
 
Manufacturer Narrative
(b)(4).Should the information be provided later, a supplemental medwatch will be sent.Additional information received: rep reported trigger of device broke.The issue occurred on second firing of device.The customer reported the tr45w reload was loaded into device, no seamguard was used, and a new ats45 device was used to complete case.The analysis results found that the ats45 device was received with the firing mechanism damaged and with a white reload present.The reload was received partially fired and with the lockout spring damaged.The damage to the reload lockout spring is consistent with damage observed when the firing cycle is started, interrupted, released, and restarted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lock out.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.No functional test could be performed due to the condition of the device.The device was disassembled to verify the condition of the internal components and the pinion axel support was found broken.While no conclusion could be reach on what cause the firing mechanism to fail, it is possible that the device was attempted to fire on thicker tissue then indicated or attempted to fire through a locked reload in previous firings causing an increase of the internal forces resulting in the component yielding.It should be noted that a 100% inspections takes place during manufacturing to ensure the device meets the require specifications; in addition, a sample of the batch is inspected at (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 0096 9
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 0096 9
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key4321128
MDR Text Key17582150
Report Number3005075853-2014-08558
Device Sequence Number1
Product Code GDW
Combination Product (y/n)N
PMA/PMN Number
K020779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/13/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/30/2019
Device Catalogue NumberATS45
Device Lot NumberL4EV8T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/01/2014
Initial Date FDA Received12/11/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TR45W
-
-