(b)(4).Should the information be provided later, a supplemental medwatch will be sent.Additional information received: rep reported trigger of device broke.The issue occurred on second firing of device.The customer reported the tr45w reload was loaded into device, no seamguard was used, and a new ats45 device was used to complete case.The analysis results found that the ats45 device was received with the firing mechanism damaged and with a white reload present.The reload was received partially fired and with the lockout spring damaged.The damage to the reload lockout spring is consistent with damage observed when the firing cycle is started, interrupted, released, and restarted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lock out.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.No functional test could be performed due to the condition of the device.The device was disassembled to verify the condition of the internal components and the pinion axel support was found broken.While no conclusion could be reach on what cause the firing mechanism to fail, it is possible that the device was attempted to fire on thicker tissue then indicated or attempted to fire through a locked reload in previous firings causing an increase of the internal forces resulting in the component yielding.It should be noted that a 100% inspections takes place during manufacturing to ensure the device meets the require specifications; in addition, a sample of the batch is inspected at (b)(4).
|