• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALERE SAN DIEGO, INC. ALERE HCG COMBO CASSETTE (25); HCG/PREGNANCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO, INC. ALERE HCG COMBO CASSETTE (25); HCG/PREGNANCY Back to Search Results
Model Number FHC-202
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/07/2014
Event Type  malfunction  
Event Description
Caller alleged false negative hcg results on two lots.Results as follows: one patient sample gave a false negative hcg result on two lots.The customer did not believe the test result, so they had a quantitative serum test done that came out to approx 35miu/ml.The same urine sample was ran on their chemistry analyzer and received 37miu/ml, though the customer states urine is not approved for use on their chemistry analyzer.They then tested a third lot of the same alere hcg test and it showed a positive result.The customer also prepared hcg dilutions for a sample approx 25miu and it showed negative on the initial two lots and positive on the most recent third lot of the alere hcg test.The customer suspects that the patient had an ectopic pregnancy or had miscarried, but could not provide any further details.No treatments or procedures were performed on the basis of the false negative results.The patient was treated according to the serum quantitative result.Additional information: external controls were previously done with alere brand controls and were ok; internal control was showing up; sample was fresh and at room temperature for initial testing; no urinalysis was performed but the urine appeared normal and clear; no sample remains to return; test was read at 3 minutes using a timer.Refer to mdr #2027969-201401060 for second lot number.
 
Manufacturer Narrative
Customer's complaint was not replicated in-house with retention devices.Retention devices were tested with cutoff hcg urine control and 3 high level hcg urine controls.All results were positive at read time.Manufacturing batch record review did not uncover any abnormalities.Patient's serum sample was quantified at approx 35 miu/ml.Hcg levels in serum are usually higher than in urine.Hcg levels in urine may have been determined without patient specimen in-house analysis.Product deficiency was not established.This issue will be tracked and trended.Corrective action is not required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALERE HCG COMBO CASSETTE (25)
Type of Device
HCG/PREGNANCY
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
san diego CA
Manufacturer Contact
ya-ling king
9975 summers ridge rd
san diego, CA 92121
8588052084
MDR Report Key4322860
MDR Text Key18545230
Report Number2027969-2014-01059
Device Sequence Number1
Product Code JHI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K993065
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/07/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFHC-202
Device Lot NumberHCG3120156
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/07/2014
Initial Date FDA Received12/03/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-