|
Catalog Number RBY2C2060 |
Device Problem
Difficult or Delayed Positioning (1157)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/12/2014 |
Event Type
malfunction
|
Manufacturer Narrative
|
Result: the pet lock was intact with the pusher assembly; the pusher assembly was kinked in multiple locations, and the coil was intact with the pusher assembly.The outer diameter of the coil and distal detachment tip (ddt) were measured and found to be within specification.Conclusion: the complaint has been evaluated.The compliant indicated that the patient was undergoing a coil embolization procedure using a ruby coil.During the procedure, the physician experienced difficultly forming two coils (7x40 and 8x25) due to high blood flow.Both coils were removed for later use in the procedure; however one of the coils (7x40) would not pull back completely into the sheath.The physician was able to use a pod8 successfully during the coil embolization.As the physician attempted to reload one of the previously used coils (8x25), the physician decided not to use it due to the blood and stickiness.Another ruby coil (20x60) was used; however, would not load into the px slim delivery microcatheter.The microcatheter was checked for kinks, it was flushed, and a guide wire was passed through it; however, no issue was noted.Evaluation of the returned device revealed that the pusher assembly had multiple kinks in the hypo-tube; however, the 7x40 coil involved in this complaint was not returned for evaluation as it was disposed of.The noted damage along the hypo-tube typically occurs from packaging the device to be returned for evaluation.Due to the damage in the pusher assembly penumbra's investigators were unable to confirm the issue in the complaint.The outer diameter of the coil and distal detachment tip (ddt) were measured and found to be within specification.These devices are 100% functionally tested and visually inspected during process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Event Description
|
The patient was undergoing a coil embolization procedure using a ruby coil.During the procedure, the physician experienced difficultly forming two ruby coils (7x40 and 8x25) due to high blood flow.Both coils were removed for a later use in the procedure; however one of the coils (7x40) would not pull back completely into the sheath.The physician was able to use a pod8 successfully.As the physician attempted to reload one of the previous coils (8x25), the physician decided not to use it due to the blood and stickiness.Another ruby coil (20x60) was used; however, would not load into the px slim delivery microcatheter.The microcatheter was checked for kinks, was flushed, and a guide wire was passed through it; however, no issue was noted.The procedure continued using different ruby coils through the same microcatheter.There was no report of an adverse effect on the patient.
|
|
Manufacturer Narrative
|
Conclusion: the device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.This mdr is associated with mdr 3005168196-2014-00859.
|
|
Search Alerts/Recalls
|
|
|