Model Number MICL 12.6
Device Problems Bent (1059); Torn Material (3024)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/19/2014
Event Type  malfunction  
Event Description
The reporter stated the surgeon was inserting a 12.6mm micl 12.6 implantable collamer lens and noted one haptic was bent/damaged.There was no patient contact and the backup lens was implanted.
 
Manufacturer Narrative
(b)(4): evaluation: method - work order search.Results - a lens work order search was performed and no similar complaints were found within the work order.Visual inspection of the returned product found one haptic torn, with a piece torn off and missing.The lens was returned in liquid.Conclusions - (no conclusion can be drawn): based on the complaint history, work order search and the evaluation of the returned product, a specific root cause of the event could not be determined.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIAN ICL (IMPLANTABLE COLLAMER LENS)
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
haupstrasse 104
nidau, CH-25 60
SZ  CH-2560
Manufacturer (Section G)
STAAR SURGICAL COMPANY
haupstrasse 104
nidau, CH-2 560
SZ   CH-2560
Manufacturer Contact
althea watson
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key4327318
MDR Text Key16178386
Report Number2023826-2014-01108
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 11/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date04/30/2015
Device Model NumberMICL 12.6
Other Device ID NumberDIOPTER -15.5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/25/2014
Initial Date FDA Received12/12/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/27/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
INJECTOR MODEL AND LOT NUMBER UNK; CARTRIDGE MODEL AND LOT NUMBER UNK
-
-