Date received: (b)(6), 2014.(b)(6).Product evaluation: when received for analysis, there was evidence of biologicals from visual inspection and reactivity with hydrogen peroxide.The hub and a portion of the inner shaft were separate from the rest of the device, and the tip of the bur was recessed into the outer tube.The inner shaft was broken 0.62" from the distal face of the inner hub which would have resulted in the reported malfunction.The break point corresponds to the proximal end of the outer tube in the front hub.When viewed under magnification, there were striations at the break point indicating metal on metal contact and an uneven break.The information indicates excess pressure was applied during use which caused deformation of the locking area, which then caused the inner shaft and outer tube to rub together until the inner shaft broke.(b)(4).
|