The customer has reported that they have received questionable results for a total of four patients tested for roche cardiac d-dimer (ddi) on a cobas h232 meter when compared to the d-dimer method on a siemens ca 1500 instrument.Of the four patients, two had results that were erroneous.The first patient had a sample tested for ddi and it resulted as 0.41 ug/ml on (b)(6) 2014.The first patient also had another sample tested on the siemens ca 1500 instrument and it resulted as 1.3 mg/l feu for d- dimer on (b)(6) 2014.The second patient, a male born on (b)(6) 1943, had a sample tested for ddi and it resulted as 0.34 ug/ml on (b)(6) 2014.The second patient also had another sample tested on the siemens ca 1500 instrument and it resulted as 0.8 mg/l feu for d-dimer on (b)(6) 2014.The patients were not adversely affected.The serial number of the cobas h232 meter is ks0206555.This meter is not sold in the united states.
|