• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. GYNECARE THERMACHOICE III UTERINE BALLOON THERAPY; DEVICE, THERMAL ABLATION, ENDOMETRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. GYNECARE THERMACHOICE III UTERINE BALLOON THERAPY; DEVICE, THERMAL ABLATION, ENDOMETRIAL Back to Search Results
Catalog Number TC003
Device Problem Balloon leak(s) (1052)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/21/2014
Event Type  malfunction  
Event Description
It was reported that the patient underwent thermal ablation on (b)(6) 2014.During the procedure, d5w was instilled and the heat cycle was activated.Less than a minute after inserting the catheter and attempting to inflate the balloon, it was discovered that the balloon did not inflate.The catheter was removed and a hole was noted at the end of the balloon.The procedure was completed with a like device.There was no adverse patient consequence reported.
 
Manufacturer Narrative
(b)(4).Conclusion: actual device was returned for evaluation.Visual and functional inspections were performed.Upon functional evaluation, one pinhole was found on the balloon in zone b.In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE THERMACHOICE III UTERINE BALLOON THERAPY
Type of Device
DEVICE, THERMAL ABLATION, ENDOMETRIAL
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 015
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125, co
l salvacar
ciudad juarez 3260 4
MX   32604
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key4330401
MDR Text Key21237256
Report Number2210968-2014-16729
Device Sequence Number1
Product Code MNB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberTC003
Device Lot NumberHJMG03
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/21/2014
Initial Date FDA Received12/15/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-