• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY ULTRAFLEX? TRACHEOBRONCHIAL; PROSTHESIS, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY ULTRAFLEX? TRACHEOBRONCHIAL; PROSTHESIS, ESOPHAGEAL Back to Search Results
Model Number M00564790
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that an ultraflex tracheobronchial stent was used in the right trachea during a stent placement procedure performed on an unknown date.During the procedure, the stent failed to deploy and the catheter was kinked.The stent was removed from the patient and the procedure was completed with another ultraflex tracheobronchial stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good and stable.
 
Manufacturer Narrative
Investigation results of stent partially deployed investigation results: a visual examination of the returned device found that the stent was partially deployed by 15mm.It was noted that the catheter shaft was kinked both proximal and distal to the stent.No issues were noted with the crochet stitches from the point of release from the stent.During analysis it was possible to retract the deployment suture and fully deploy the stent with some bending of the shaft noted.The noted issue is likely due to procedural or anatomical factors encountered during the procedure such as tortuous anatomy or maneuvering of the device.Therefore, the most probable root cause for this complaint is operational context.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications.A labeling review was performed and, from the information available, this device was used per the directions for use (dfu) / product label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX? TRACHEOBRONCHIAL
Type of Device
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
ballybrit business park
galway
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
ballybrit business park
galway
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key4332821
MDR Text Key5219499
Report Number3005099803-2014-03930
Device Sequence Number1
Product Code ESW
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
K012883
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/25/2015
Device Model NumberM00564790
Device Catalogue Number6479
Device Lot Number16193161
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/25/2014
Initial Date FDA Received12/16/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/03/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-