Investigation results of stent partially deployed investigation results: a visual examination of the returned device found that the stent was partially deployed by 15mm.It was noted that the catheter shaft was kinked both proximal and distal to the stent.No issues were noted with the crochet stitches from the point of release from the stent.During analysis it was possible to retract the deployment suture and fully deploy the stent with some bending of the shaft noted.The noted issue is likely due to procedural or anatomical factors encountered during the procedure such as tortuous anatomy or maneuvering of the device.Therefore, the most probable root cause for this complaint is operational context.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications.A labeling review was performed and, from the information available, this device was used per the directions for use (dfu) / product label.
|