Reportable based on device analysis completed on 21nov2014.It was reported that the rotawire was kinked.The target lesion was located in the left circumflex artery.A rotawire¿ and wireclip¿ torquer was selected to treat the target lesion.During the procedure, while introducing the device inside the patient, a kink was noted in the mid segment of the rotawire.The procedure was completed with a different device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed that the spring tip of the rotawire was partially detached.
|