Model Number H802228240022
Device Problem Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/25/2014
Event Type  malfunction  
Event Description
Reportable based on device analysis completed on 21nov2014.It was reported that the rotawire was kinked.The target lesion was located in the left circumflex artery.A rotawire¿ and wireclip¿ torquer was selected to treat the target lesion.During the procedure, while introducing the device inside the patient, a kink was noted in the mid segment of the rotawire.The procedure was completed with a different device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed that the spring tip of the rotawire was partially detached.
 
Manufacturer Narrative
(b)(4).The device was returned for analysis.The visual inspection was performed and reveled a kink along the body and the spring tip was partially detached.All the outer diameter measurements are within the specifications.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAWIRE? AND WIRECLIP? TORQUER
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS  
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key4334224
MDR Text Key5140849
Report Number2134265-2014-07613
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 11/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/09/2016
Device Model NumberH802228240022
Device Catalogue Number22824-002
Device Lot Number16803523
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/21/2014
Initial Date FDA Received12/16/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/11/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial