• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR CP; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR CP; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number ADVIA CENTAUR CP
Device Problem Low Test Results (2458)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/25/2014
Event Type  malfunction  
Event Description
A discordant, falsely low troponin result was obtained on one patient sample on an advia centaur cp instrument.The discordant result was not reported to the physician(s).The sample was repeated twice on the same instrument, resulting higher.The corrected result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely low troponin result.
 
Manufacturer Narrative
The customer contacted the siemens customer care center (ccc).The sample resulted as expected upon repeat testing on the instrument.The cause of the discordant, falsely low troponin result is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR CP
Type of Device
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
STRATEC BIOMEDICAL SWITZERLAND AG
registration #: 3008494306
gewerbestrasse 6
neuhausen am rheinfall, 8212
SZ   8212
Manufacturer Contact
cassandra kocsis
511 benedict avenue
tarrytown 10591
9145242687
MDR Report Key4338473
MDR Text Key5178877
Report Number2432235-2014-00716
Device Sequence Number1
Product Code MOI
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K971418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 11/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberADVIA CENTAUR CP
Device Catalogue Number086-A002-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/25/2014
Initial Date FDA Received12/17/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-