Customer reported a treatment was in cycle six collect phase when an intermittent noise was heard to be coming from the top of the centrifuge bowl.Customer reported a blood leak alarm then occurred, with condensation appearing under the centrifuge lid.Customer stated the treatment was aborted at that time and the bowl was immediately clamped off.Customer stated she consulted with the physician assistant, and then completed a manual return of most of the blood in the kit to the patient, except for the volume in the bowl and in the photoactivation module.Customer estimated about 150 ml of blood was not returned to the patient.Customer stated when the inside of the centrifuge was wiped with gauze, only one drop of blood was observed on the gauze.Customer reported the centrifuge drain bag was empty.Customer reported patient is stable.Customer stated patient reported feeling a little light-headed immediately after the leak occurred, but the patient felt normal within ten minutes.Customer indicated no field service is needed.No service order was generated.Kit was returned for investigation.
|