• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTERVASCULAR HEMAGARD KNITTED GRAFT; VASCULAR GRAFT PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTERVASCULAR HEMAGARD KNITTED GRAFT; VASCULAR GRAFT PROSTHESIS Back to Search Results
Model Number HGK0008-20
Device Problems Improper or Incorrect Procedure or Method (2017); Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem Blood Loss (2597)
Event Date 11/09/2014
Event Type  malfunction  
Event Description
During an ascending aorta aneurysm repair performed on (b)(6) 2014, the involved graft was used as a cannulation for blood transmission.The graft was used as a cannulation for blood transmission.The graft was reported to leak during the procedure.The graft clotted after a wait and was not replaced.No pt injury was reported.
 
Manufacturer Narrative
A review of the device history records, including collagen coating records, indicated that the graft was processed and inspected according to procedures and no anomaly was found.Specifically, the review of the water permeability testing records of products coated on the same day and under the same conditions as the complaint device indicated values will within product specifications.One retention sample coated in the same period and under the same conditions as the complaint device underwent water permeability testing at 120mmhg as per iso 7198.The test result indicated a value well within product specifications.Please note that the device was not used in an approved indication.No conclusion can be drawn.However, all available info and the product testing performed would tend to indicate that the device was not defective.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMAGARD KNITTED GRAFT
Type of Device
VASCULAR GRAFT PROSTHESIS
Manufacturer (Section D)
INTERVASCULAR
zi athelia 1
la iotat cedex 1360 5
FR  13605
Manufacturer Contact
pascal de framond
zi athelia 1
la ciotat cedex 13605
FR   13605
42084646
MDR Report Key4340538
MDR Text Key18031049
Report Number1640201-2014-00035
Device Sequence Number1
Product Code DSY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K964625
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 11/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2019
Device Model NumberHGK0008-20
Device Catalogue NumberHGK0008-20
Device Lot Number14F19
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/12/2014
Initial Date FDA Received12/09/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-