It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the potassium (k+) number kept drifting higher during the case on the blood parameter monitor (bpm).The device was not changed out, as they just continued to use the monitor.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.Per the clinical review on (b)(6) 2014: the perfusionist (ccp) stated that k+ drift is an intermittent issue (not every case), but in this case was especially inaccurate.After cpb was initiated and cardioplegic arrest occurred, an in-vivo calibration was performed about five minutes after the completion of the cardioplegia dose.After the results were entered into the bpm, the k+ measure would drift upward.According to the ccp, an in-vivo recalibration was performed every 20 minutes and the k+ would drift between each re-calibration.During rewarming of the patient (later in cpb), the bpm measured the k+ as 8.0 mmol/l and the laboratory analyzed value was 5.7 mmol/l.According to the ccp, there were no issues with the other bpm shunt/bpm measures.The case was completed successfully, without delay and without associated blood loss.According to the ccp, more frequent laboratory analysis was needed (every 20 minutes) in this case.There was no harm observed.
|