The device history record (dhr) was reviewed and no deviations related to this failure mode were found.There are no non-conformances related to the reported issue.The product sample returned consisted of one 14.5 fr palindrome catheter, 19 cm implant length, 36 cm with signs of use.After a visual inspection was performed, a hole was found on the joint of the catheter below the hub.During functional testing (underwater test), bubbles were detected coming out below the hub, from the lumen which corresponds to the arterial extension.The lumen related to the venous extension did not show bubbles during the test.As per the instructions for use, the customer has to perform a visual inspection before using the device.Do not use the catheter if it appears damaged or defective.The catheter tubing can tear when subjected to excessive force or rough edges.Inspect the catheter frequently for nicks, scrapes or cuts which could impair its performance.The catheter functioned as intended for 2 years and 4 months.The issue was noticed during use.The most probable root cause could be due to damage during the medical procedure.The leak can be due to excessive force or inappropriate use of cleaning agents.The lot involved on this event was manufactured on (b)(6) 2012, before the implementation date of actions related to a corrective and preventative action.It was defined that this event is being handled through this capa and no additional actions are required.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling) are in place to prevent nonconforming product from leaving the manufacturing operations.This complaint will be used for tracking and trending purposes.
|