Catalog Number 10309478
Device Problem Patient Data Problem (3197)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/21/2014
Event Type  malfunction  
Event Description
Customer reported that instrument displayed erroneous patient sample id when they scanned the label on a tube.Customer also reported that patient sample id was missing 1 digit and sometimes it was replaced by another number or a letter and then randomly.There was no report of injury due to this event.
 
Manufacturer Narrative
Siemens customer service engineer (cse) indicated that internal barcode reader had been replaced by the in-house service technician.Cse also indicated that there have been no occurrences of a digit changing to another since replacing the barcode reader.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK ATLAS
Type of Device
CT ATLAS
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown 10591
Manufacturer (Section G)
SPARTON MEDICAL
22740 lunn road
cleveland OH 44149
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key4342906
MDR Text Key5202924
Report Number1217157-2014-00194
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K932674
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Medical Technologist
Device Catalogue Number10309478
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/24/2014
Initial Date FDA Received12/18/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-