Result: the ruby coil was detached from the pusher assembly.The outer diameter of the coil, and proximal constraint sphere were measured and found to be within specification.Conclusion: the complaint has been evaluated.The complaint indicated that the ruby coil unintentionally detached inside a 0.027 microcatheter (a non-penumbra device).Evaluation of the returned device revealed that the ruby coil was detached from the pusher assembly and that the ruby coil was intact.A non-penumbra wire was returned; however, the pusher assembly for the ruby coil and the microcatheter used in the procedure were not returned for a complete evaluation.Without the return of the pusher assembly, we are unable to determine and/or confirm if a premature detachment occurred.It is also unknown if there was any damage to the 0.027 microcatheter prior to the premature detachment.The root cause of this complaint cannot be determined.These devices are 100% functional tested during process inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|