• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRODUCTOS PARA EL CUIDADO DE LA SALUD 2-WAY 5CC ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRODUCTOS PARA EL CUIDADO DE LA SALUD 2-WAY 5CC ALL-SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 165816
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported that the catheter developed a ridge upon deflation.The ridge created a delay in removal.
 
Manufacturer Narrative
The sample was not returned for eval, however, a picture was attached to the investigation file where a catheter can be observed showing a balloon cuff roll.The lot number is unk, therefore, the device history record could not be reviewed.There are no changes in the manufacturing process that would have contributed adversely with the reported problem.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2-WAY 5CC ALL-SILICONE FOLEY CATHETER
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD
nogales, sonora
MX 
Manufacturer (Section G)
PRODUCTOS PARA EL CUIDADO DE LA SALUD
km 7 carretera internacional
nogales, sonora 8562 1
MX   85621
Manufacturer Contact
christy lewis
8195 industrial blvd.
covington, GA 30014
7707846100
MDR Report Key4349337
MDR Text Key5178012
Report Number1018233-2014-00343
Device Sequence Number1
Product Code EZL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040504
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number165816
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/19/2014
Initial Date FDA Received12/12/2014
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-