• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRIS INTERNATIONAL IQ200; AUTOMATED URINE MICROSCOPY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRIS INTERNATIONAL IQ200; AUTOMATED URINE MICROSCOPY ANALYZER Back to Search Results
Catalog Number 700-3375
Device Problems False Negative Result (1225); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/14/2014
Event Type  malfunction  
Event Description
Customer reported iq200 not reading dilution barcodes.
 
Manufacturer Narrative
Customer reported an ongoing intermittent issue with the iq200 misreading dilution labels and erroneously setting dilution ratio 1:1.There was no change to patient management, however, erroneous results were reported out to the physician.Iris field service engineer (fse) replaced the barcode reader and the system was operational.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IQ200
Type of Device
AUTOMATED URINE MICROSCOPY ANALYZER
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer Contact
sudha gupta
9172 eton avenue
chatsworth, CA 91311
8185277272
MDR Report Key4349777
MDR Text Key5200300
Report Number2023446-2014-00185
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022774
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number700-3375
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2014
Initial Date FDA Received12/05/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-