• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS FT4, FREE THYROXINE; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 06437281190
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/05/2014
Event Type  malfunction  
Event Description
The customer received a questionable free thyroxine (ft4) result from the cobas 8000 e602 analyzer compared to the result from an architect analyzer.The result from the cobas e602 was 53 pmol/l.On (b)(6) 2014, the same sample was tested on an abbott architect analyzer and the result was 15.81 pmol/l.The erroneous result was reported outside of the laboratory.The patient was not adversely affected.The preliminary investigation could not determine a specific root cause.Further investigation of the sample for possible interference was not possible as no sample was available.Based on the provided calibration and qc data, a general reagent issue was not suspected.
 
Manufacturer Narrative
This event occurred in (b)(6).
 
Manufacturer Narrative
A specific root cause could not be identified.Based on the provided calibration and qc information, a general reagent issue could be excluded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 6830 5
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key4350760
MDR Text Key5144361
Report Number1823260-2014-10270
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 01/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number06437281190
Device Lot Number179279
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Initial Date Manufacturer Received 12/16/2014
Initial Date FDA Received12/22/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/23/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-