It was reported on 01/19/2015 that the patient¿s lead and generator were explanted on (b)(6) 2015 due to infection.The patient was not re-implanted at this time.The lead and generator were returned for analysis on 01/21/2015.Product analysis was completed for the lead on 02/09/2015.Other than typical wear and explant related observations, no anomalies were identified in the returned lead portions.Since the electrode array portion was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.Product analysis was completed and approved for the generator on 02/17/2015.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.The battery, 3.034 volts as measured during completion of test parameter of the final electrical test, shows an ifi=no condition.There were no performance or any other type of adverse conditions found with the pulse generator.
|