Model Number 722003
Device Problem Bent (1059)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Philips received a complaint from a customer that the fluroscopy failed during the study due to a bent footswitch.
 
Manufacturer Narrative
(b)(4).Philips received a complaint from a customer that the fluoroscopy failed during the study due to a bent footswitch.The field service engineer (fse)'s visual review of the system showed the footswitch was bent and not working properly.Philips conducted its investigation based on information received.Based on available information, the cause could be related to the use of anti-fatigue mats under the footswitches since the footswitch was designed to be used on a solid surface.After adjustment of the footswitch, the reported problem with the system was solved.Philips is working on adding a metal plate under the footswitch to prevent bending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10
Type of Device
SYSTEM, X RAY SYSTEM, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5680 DA
NL  5680 DA
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5680 DA
NL   5680 DA
Manufacturer Contact
deborah thurston
3000 minuteman rd
andover, MA 01810
9786592010
MDR Report Key4355787
MDR Text Key5222445
Report Number3003768277-2014-00134
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031333
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Health Professional
Type of Report Initial
Report Date 10/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number722003
Device Catalogue Number722003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/21/2014
Initial Date FDA Received12/17/2014
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-